91-9873668449, Landline :+91-120-4114328
KUK GROUP


NOMEX Dosemeter

 

  • CE marked, class IIb certified diagnostic dosimetry system, fully compliant with IEC 61674 for acceptance testing and quality control measurements on RAD, FLU, DENT, CT and MAM X-ray modalities
  • Field-class electrometer acc. to IEC 60731 for air kerma measurements in radiotherapy and orthovoltage therapy
  • Outstanding accuracy (± 2%) and excellent resolution (20 fA)
  • Connection of semiconductor detectors or ion chambers or of the NOMEX® Multimeter
  • Combined with the NOMEX Multimeter and/or NOMEX mAs all relevant QC parameters can be captured in one single shot
  • Automatic air density correction when using ion chambers
  • Waveform display and storage of last 100 measurements
  • Allows data and waveform export as XLS, XML or CSV file via USB or Bluetooth®
  • Operation menu in multiple languages (Chin./Eng./Fren./Ger./Ital./Jap./Port./Russ./Span.)
  • Convenient firmware updates on-site

  • 100% precision

    Outstanding Precision and Excellent Resolution: The NOMEX® Dosemeter comes electrically calibrated with an intrinsic error of max. ± 2%. Measuring results of highest accuracy and of a superior reading resolution (20 fA) are guaranteed.

  • 100% ease of use

    Connect the detector of choice (usage of already available PTW diagnostic detectors!), position the detector and start the measurement.

  • Sophisticated Warning and Error Management

    NOMEX® displays when a measurement is inaccurate and suggests what to change within the settings or the measurement setup.

  • Time-Saving, Straightforward and Cost-Effective Service and Calibration Workflows

    Firmware updates can be easily performed via USB connection to a PC. No need to return the device to the manufacturer. PTW devices are worldwide well-known for their reliability, reproducibility and long-term stability. A recalibration is recommended only every 2 years.

  • Individual sets for every X-ray modality available: Different NOMEX sets with or without the NOMEX Multimeter and each with or without mAs option are available!

  • Service contract for NOMEX® Dosemeter (Bronze, Silver, Gold)

Packages

Each package is available with or without mAs option:

  • NOMEX Dosemeter
  • NOMEX set all-in R/F/MAM/DENT/CT
  • NOMEX set R/F/MAM
  • NOMEX set R/F
  • NOMEX set MAM
  • NOMEX set DENT
  • NOMEX set CT

Detectors


Semiconductors detectors

  • NOMEX® Multimeter
  • R/F/D detector
  • MAM detector
  • SL Survey Detector

Ionization chambers

  • SFD diagnostic chamber 75 cm³
  • SFD MAM chamber 6 cm³
  • CT ion chamber, 100 mm long
  • CT ion chamber, 300 mm long
  • Semiflex 3D, 0.07cm³ chamber
  • Farmer type chamber 0.6 cm³, waterproof
  • Soft X-ray chamber 0.2 cm³
  • NOMEX mAs (invasive measurement)
Current clamp for NOMEX mAs (non-invasive measurement)

Measuring Ranges

Current Resolution
2 pA ... 23 µA, < ± 2% ± 20 fA
Charge Resolution
2 pC ... 230 mC, < ± 2% ± 10 fC
Irradiation time
1 ms ... 298 h, ± 1 ms
Pulses
0 ...99999, ± 1 pulse
Frequency
(0.2 ... 500) Hz
Temperature
(10 ... 40) °C ± 0.1 °C
Pressure
(750 ... 1100) hPa ± 0.1 hPa
Dimensions (HxWxD)
160 mm x 150 mm x 40 mm
Weight
550 g incl. rechargeable batteries
Interfaces
USB V2.0 High-Speed, Bluetooth® 2.1 + EDR


NOMEX Multimeter


R/F/D detector (type 60004)

Air kerma rate
40 nGy/s ... 460 mGy/s
Air kerma
40 nGy ... 4.6 kGy
Resolution
0.4 nGy/s
Energy dependence
≤± 5%

MAM detector (type 60005)

Air kerma rate
130 nGy/s ... 1.5 Gy/s
Air kerma
130 nGy ... 15 kGy
Resolution
1.3 nGy
Energy dependence
≤± 5%

CT detector (type 30009)

Air kerma rate
140 µGy*cm/s ... 1.6 kGy*cm/s
Air kerma
140 µGy*cm ... 16 MGy*cm
Resolution
1.4 µGy*cm/s
Energy dependence
≤± 5%